Radiation protection value to the operator from augmented reality smart glasses in interventional fluoroscopy procedures using phantoms

22/11/2019

0 Comments

 
Author(s) Dorey S.; Gray L.; Tootell A.; Higgins R.; Hogg P. et al.
Source Radiography; Nov 2019; vol. 25 (no. 4); p. 301-307
DOI 10.1016/j.radi.2019.02.008
Introduction: Smart glasses can be adapted to display radiographic images to allow clinician's gaze not to be directionally fixed or predetermined by computer monitor location. This study presents an analysis of eye lens dose during interventional fluoroscopy guided procedures, comparing fixed monitor positions against the use of smart glasses.

Method(s): Using a head phantom (simulating the clinician), thermoluminescent dosimeters and lead shielded glasses, the dose to the eye was measured for different head 'rotations and tilts' for: gaze directed towards the main scattering source (patient/primary beam) to represent potential gaze direction if smart glasses are used; gaze directed to a range of potential computer monitor positions. An anthropomorphic pelvis phantom was utilised to simulate the patient. Accumulated dose rates (muGy s1) from five 10-second exposures at 75 kV 25.2 mAs were recorded.

Result(s): An average DAP reading of 758.84 cGy cm2 was measured during each 10 second exposure. Whilst wearing lead shielded glasses a 6.10 - fold reduction in dose rate to the lens is possible (p < 0.05). Influence of the direction of gaze by the clinician demonstrated a wide range of dose rate reduction from 3.13% (p = 0.16) to 143.69% (p < 0.05) when the clinician's gaze was towards the main scattering source. Increased dose rate to the clinician's eyes was received despite wearing lead shielded glasses, as the angle of gaze moved 45degree and 90degree from 0degree.

​Conclusion(s): If the clinician's gaze is directed towards the main scattering source a potential exists for reducing eye lens dose compared with fixed location computer monitors. Introduction of lead lined smart glasses into interventional radiology may lead to improvements in patient care, reducing the need for the clinician to look away from the patient to observe a radiographic image.Copyright © 2019
0 Comments



Leave a Reply.

    The following databases were searched:
    EMBASE,  MEDLINE, PsycINFO, BNI, CINAHL, 
    to find  ELHT staff publications

    Specialties

    All
    ANAESTHETICS
    CARDIOLOGY
    COVID 19
    DERMATOLOGY
    DIABETES
    DIETETICS
    EMERGENCY CARE
    GYNAECOLOGY
    MAXILLOFACIAL
    NEONATOLOGY
    OBSTETRICS
    ONCOLOGY
    ORTHOPAEDICS
    PAEDIATRICS
    PAIN
    PHARMACY
    RADIOLOGY
    RESPIRATORY
    SURGERY
    UROLOGY
    VASCULAR SURGERY

    RSS Feed

    Archives - past 2 years

    December 2020
    November 2020
    October 2020
    September 2020
    July 2020
    June 2020
    May 2020
    April 2020
    March 2020
    February 2020
    January 2020
    December 2019
    November 2019
    October 2019
    September 2019
    August 2019
    January 2019
    December 2018

Learning Centre Library
Royal Blackburn Teaching Hospital
library.blackburn@elht.nhs.uk
01254 734312 or Ext 84312

​Find us

Staffed Opening Hours
Mon      08:30-16:30
Tue       08:30-16:30
Wed     08:30-16:30
Thu      08:30-16:30
Fri         08:30-16:00
Mackenzie Library
Burnley General Teaching Hospital
library.burnley@elht.nhs.uk
01282 803114 or Ext. 13114

Find us

Staffed Opening Hours
Mon      08:30-16:00
Tue       08:30-16:00
Wed     08:30-16:00
Thu       08:30-16:00
Fri         Unstaffed

FAQS
​Library Catalogue
Library @lerts
​Your Specialty or Professional Group
Membership Terms & Conditions
Privacy Policy
Our Performance
​Your Feedback
​Specialty Journals

Picture
Picture
24/7 access to the library is available - please see library staff
Picture
  • Home
  • Meet the Team
  • Library Facilities
  • Education, Research and Innovation